Key result
Braile Inovare transcatheter aortic valve-in-valve implantation shows feasibility with 0% operative mortality in high-risk patients.
Why the study?
Aortic valve replacement for bioprosthesis dysfunction involves considerable risk and high mortality, making minimally invasive transcatheter valve-in-valve implantation a potential alternative to reduce morbidity and mortality.
Is transcatheter aortic valve-in-valve implantation using the Braile Inovare prosthesis safe and effective in patients with aortic bioprosthesis dysfunction?
Observational (n=14)
Is transcatheter aortic valve-in-valve implantation using the Braile Inovare prosthesis safe and effective in patients with aortic bioprosthesis dysfunction?
Answers are grounded in this paper and cite their sources.
Transcatheter aortic valve-in-valve implantation using the Braile Inovare prosthesis is a feasible and safe alternative for high-risk patients with aortic bioprosthesis dysfunction.
OBJECTIVE: Aortic valve replacement for bioprosthesis dysfunction is a procedure involving considerable risk. In some cases, mortality is high and may contraindicate the procedure. Minimally invasive transcatheter aortic "valve-in-valve" implant appears to be an alternative, reducing morbidity and mortality. The objective is to evaluate aortic valve-in-valve procedure using Braile Inovare prosthesis. METHODS: The Braile Inovare prosthesis, transcatheter, expandable balloon, was used in 14 cases. Average EuroSCORE was 42.9%. All patients had double aortic bioprosthesis dysfunction. Procedures were performed in a surgical hybrid environment under echocardiographic and fluoroscopic guidance. Using left minithoracotomy prostheses were implanted through the ventricular apex under high-frequency ventricular pacing. Serial clinical and echocardiographic controls were performed. Follow-up ranged 1-30 months. RESULTS: Correct prosthetic deployment was obtained in all cases. There was no conversion. There was no operative mortality. The 30-day mortality was 14.3% (two cases). Ejection fraction increased significantly after the 7th postoperative day. Aortic gradient significantly reduced. The residual aortic regurgitation was not present. There were no vascular complications or complete atrioventricular block. CONCLUSION: The transcatheter "valve-in-valve" procedure for bioprosthesis dysfunction is safe with low morbidity. This possibility may change prosthesis choice during the first aortic valve replacement, favoring bioprostheses.
No takes yet. Share an insight, caveat, or question.
Gaia et al. (2012) conducted an observational in double aortic bioprosthesis dysfunction (n=14). Transcatheter aortic valve-in-valve implantation using Braile Inovare prosthesis was evaluated on 30-day mortality. Transcatheter aortic valve-in-valve implantation using the Braile Inovare prosthesis was safe and feasible in 14 high-risk patients, with 0% operative mortality and 14.3% 30-day mortality.