Answers are grounded in this paper and cite their sources.
High-resolution, long-term measurements of ice nucleating particles (INPs) have been impeded by complex instrumentation that requires a trained on-site technician to operate or analyze offline. We have significantly updated the well-characterized continuous flow diffusion chamber (CFDC) instrument to run autonomously with minimal in-person handling and easy remote access. This new CFDC, the CFDC-Ice Activation Spectrometer (CFDC-IAS) was deployed for four months (October 2020–January 2021) at the mountain-top Storm Peak Laboratory site in Colorado and provided 5-minute resolution measurements daily at target temperatures of −20, −25, and −30 °C. Concentrations of INPs across all temperatures had a median value of 6 per standard liter (sL−1), and a mean of 10 sL−1 with a range of ~0–470 sL−1.
No takes yet. Share an insight, caveat, or question.
Hodshire et al. (2022) studied this question.