Key result
Modular LP and S-ICD system achieves ~100% wireless communication success in preclinical testing.
Why the study?
Subcutaneous ICDs do not deliver antibradycardia or antitachycardia pacing, and no commercially available coordinated leadless option existed for patients needing both defibrillation and pacing.
Does a novel modular cardiac rhythm management system combining a wirelessly communicating antitachycardia pacing-enabled leadless pacemaker and a subcutaneous implantable cardioverter-defibrillator provide stable long-term communication, electrical performance, and safety in a canine model?
Does a novel modular cardiac rhythm management system combining a wirelessly communicating antitachycardia pacing-enabled leadless pacemaker and a subcutaneous implantable cardioverter-defibrillator provide stable long-term communication, electrical performance, and safety in a canine model?
This preclinical canine study demonstrates that a modular cardiac rhythm management system combining a leadless pacemaker and an S-ICD maintains reliable wireless communication, stable electrical pacing parameters, and acceptable safety up to 18 months, supporting feasibility for human trials.
Answers are grounded in this paper and cite their sources.
Supports preclinical feasibility of modular LP-S-ICD systems; leaves open translation to human use.
BACKGROUND: Subcutaneous implantable cardioverter-defibrillators (S-ICDs) and leadless pacemakers (LPs) are intended to diminish transvenous lead-related complications. However, S-ICDs do not deliver antibradycardia pacing or antitachycardia pacing, and currently, there is no commercially available coordinated leadless option for patients with defibrillator and (expected) pacing needs. OBJECTIVE: We evaluated the performance, safety, and potential replacement strategies of a novel modular cardiac rhythm management (mCRM) system, a wirelessly communicating antitachycardia pacing-enabled LP and S-ICD in a preclinical model. METHODS: LP implantation was attempted in 68 canine subjects, and in 38 an S-ICD was implanted as well. Animals were evaluated serially up to 18 months. At all evaluations, communication thresholds (CTs) between the devices, LP electrical parameters, and system-related complications were assessed. Different replacement strategies were tested. RESULTS: The LP was successfully implanted in 67 of 68 (98.5%) and the concomitant S-ICD in 38 of 38 (100%). mCRM communication was successful in 1022 of 1024 evaluations (99.8%). The mean CT was 2.2 ± 0.7 V at implantation and stable afterward (18 months: 1.8 ± 0.7 V). In multivariable analysis, larger LP-to-S-ICD angle and dorsal posture were associated with higher CTs. At implantation, the mean pacing capture threshold, impedance, and R-wave amplitude were 0.3 ± 0.1 V, 898.4 ± 198.9 Ω, and 26.4 ± 8.2 mV. The mean pacing capture threshold remained stable and impedance and R-wave amplitudes were within acceptable ranges throughout (0.7 ± 0.4 V, 619.1 ± 90.6 Ω, and 20.1 ± 8.4 mV at 18 months). Different replacement strategies seem feasible. CONCLUSION: This first mCRM system demonstrated excellent performance up to 18 months in a preclinical model.
No takes yet. Share an insight, caveat, or question.
Breeman et al. (2022) studied Preclinical canine model for modular cardiac rhythm management (defibrillation and pacing needs) (n=68). Modular cardiac rhythm management (mCRM) system consisting of an antitachycardia pacing-enabled leadless pacemaker (EMPOWER) and subcutaneous implantable cardioverter-defibrillator (EMBLEM S-ICD) was evaluated on Successful wireless device-to-device (S-ICD to LP) communication. The modular LP and S-ICD system achieved 99.8% (1022/1024) wireless communication success with stable thresholds and pacing capture in a canine model.