This is the authors' abstract. We don't add key points for this paper.
Medical implants with customizable design capabilities are an important direction for high-end medical devices. An important standard in implant design is to match the mechanical properties of the implantation site. Inspired by the ‘cartilage ring-connective tissue’ structure of natural tracheal, we propose a design approach of tracheal stent composed of ‘skeleton + filling matrix’. By establishing a mapping model of structural parameters and J-type response mechanical properties, we realize a larger degree of freedom in adjusting the J-type mechanical properties. This design approach uniquely aligns the mechanical properties of the tracheal stent with those of the native trachea. We provided a thermal field-assisted rotary printing method enabling the integration of both thermoplastic and thermoset materials, which have successfully fabricated this heterogeneous structural stent as a single piece using. Our design methodology is a great inspiration for the development of medical luminal implants with designable and adjustable mechanical properties.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Tian et al. (2024) studied this question.