Key result
Real-time 2-D vector flow imaging on a phased array successfully visualized in vivo flow dynamics and estimated peak systolic velocity in the ascending aorta at 135 cm/s ± 16.9%.
Real-time 2-D vector flow imaging on a phased array is feasible and can provide quantitative and qualitative flow evaluations of complex flow dynamics.
Supports feasibility of real-time 2-D VFI in humans; leaves open clinical utility pending validation studies.
Echocardiography examination of the blood flow is currently either restricted to 1-D techniques in real-time or experimental offline 2-D methods. This paper presents an implementation of transverse oscillation for real-time 2-D vector flow imaging (VFI) on a commercial BK Ultrasound scanner. A large field-of-view (FOV) sequence for studying flow dynamics at 11 frames per second (fps) and a sequence for studying peak systolic velocities (PSVs) with a narrow FOV at 36 fps were validated. The VFI sequences were validated in a flow rig with continuous laminar parabolic flow and in a pulsating flow pump system before being tested in vivo, where measurements were obtained on two healthy volunteers. Mean PSV from 11 cycles was 155 cms-1with a precision of ±9.0% for the pulsating flow pump. In vivo, PSV estimated in the ascending aorta was 135 cms-1± 16.9% for eight cardiac cycles. Furthermore, in vivo flow dynamics of the left ventricle and in the ascending aorta were visualized. In conclusion, angle independent 2-D VFI on a phased array has been implemented in real time, and it is capable of providing quantitative and qualitative flow evaluations of both the complex and fully transverse flow.
No takes yet. Share an insight, caveat, or question.
Holbek et al. (2018) studied Healthy (n=2). Real-time 2-D vector flow imaging (VFI) on a phased array was evaluated on Peak systolic velocity (PSV) and flow dynamics visualization. Real-time 2-D vector flow imaging on a phased array successfully visualized in vivo flow dynamics and estimated peak systolic velocity in the ascending aorta at 135 cm/s ± 16.9%.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: