Answers are grounded in this paper and cite their sources.
Two-dimensional (2-D)-to-three-dimensional (3-D) registration can improve visualization which may aid minimally invasive neurointerventions. Using clinical and phantom studies, two state-of-the-art approaches to rigid registration are compared quantitatively: an intensity-based algorithm using the gradient difference similarity measure; and an iterative closest point (ICP)-based algorithm. The gradient difference approach was found to be more accurate, with an average registration accuracy of 1.7 mm for clinical data, compared to the ICP-based algorithm with an average accuracy of 2.8 mm. In phantom studies, the ICP-based algorithm proved more reliable, but with more complicated clinical data, the gradient difference algorithm was more robust. Average computation time for the ICP-based algorithm was 20 s per registration, compared with 14 min and 50 s for the gradient difference algorithm.
No takes yet. Share an insight, caveat, or question.
McLaughlin et al. (2005) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: