Key result
An SVD-based algorithm outperformed existing methods in reconstructing trunk EMG, showing better agreement with the respiratory cycle (correlation coefficient = 0.81, p<0.01) compared with TS and ICA.
Why the study?
A new method was needed to remove cardiac contamination from trunk electromyography without requiring an additional reference ECG.
Effect estimate: correlation coefficient 0.81
p-value: p=< 0.01
A novel SVD-based algorithm effectively removes cardiac interference from trunk EMG without requiring an additional reference ECG, outperforming current state-of-the-art methods.
May enhance trunk EMG quality in research; leaves open clinical utility pending patient validation.
A new algorithm based on singular value decomposition (SVD) to remove cardiac contamination from trunk electromyography (EMG) is proposed. Its performance is compared to currently available algorithms at different signal-to-noise ratios (SNRs). The algorithm is applied on individual channels. An experimental calibration curve to adjust the number of SVD components to the SNR (0–20 dB) is proposed. A synthetic dataset is generated by the combination of electrocardiography (ECG) and EMG to establish a ground truth reference for validation. The performance is compared with state-of-the-art algorithms: gating, high-pass filtering, template subtraction (TS), and independent component analysis (ICA). Its applicability on real data is investigated in an illustrative diaphragm EMG of a patient with sleep apnea. The SVD-based algorithm outperforms existing methods in reconstructing trunk EMG. It is superior to the others in the time (relative mean squared error < 15%) and frequency (shift in mean frequency < 1 Hz) domains. Its feasibility is proven on diaphragm EMG, which shows a better agreement with the respiratory cycle (correlation coefficient = 0.81, p-value < 0.01) compared with TS and ICA. Its application on real data is promising to non-obtrusively estimate respiratory effort for sleep-related breathing disorders. The algorithm is not limited to the need for additional reference ECG, increasing its applicability in clinical practice.
No takes yet. Share an insight, caveat, or question.
Peri et al. (2021) studied Cardiac interference in trunk electromyography. Singular value decomposition (SVD)-based algorithm vs. Gating, high-pass filtering, template subtraction (TS), and independent component analysis (ICA) was evaluated on Agreement with the respiratory cycle (correlation coefficient 0.81, p=< 0.01). An SVD-based algorithm outperformed existing methods in reconstructing trunk EMG, showing better agreement with the respiratory cycle (correlation coefficient = 0.81, p<0.01) compared with TS and ICA.