Answers are grounded in this paper and cite their sources.
Transcutaneous oxygen tension measurement (TcPO 2 ) is widely applied for the evaluation of chronic limb-threatening ischemia (CLTI). Nevertheless, studies that focused on the clinical value of TcPO 2 have shown varying results. We identified factors that potentially play a role in TcPO 2 measurement variation such as probe placement, probe temperature, and the use of a reference probe. In this review of the current literature, we assessed the application of these factors. A systematic search was conducted. Parameters that were assessed were probe placement, probe temperature, and mentioning and/or use of a reference probe. In total, 36 articles were eligible for analysis. In 24 (67%) studies, probes were placed on specific anatomical locations. Seven (19%) studies placed probes, regardless of the location of the ulcer, adjacent to an ischemic lesion or ulcer (perilesion). Selected temperature setting of the probe differed; in 18 (50%), a default probe temperature of 44°C was selected, and in 13 (36%), a different temperature was selected. In 31 (84%) studies, the use of a reference probe was not reported. Transcutaneous oxygen tension measurement is applied diversely in patients with CLTI. Homogeneity in TcPO 2 protocols is warranted for reliable clinical application and to compare future TcPO 2 research.
No takes yet. Share an insight, caveat, or question.
Leenstra et al. (2019) studied this question.