Answers are grounded in this paper and cite their sources.
The ADx total serum tricyclic antidepressant (TCA) fluorescence polarization immunoassay (Abbott Diagnostics) for the semi-quantitation of imipramine or amitriptyline and their respective N-demethylated metabolites in cases of TCA overdose was evaluated. The assay is linear from 75-1000 ng/mL total TCA in serum, and flaggs as "HI" all results exceeding 300 ng/mL. The within and between run precision of the assay for patient serum containing imipramine or amitriptyline and their metabolites gave CV's of less than 5.5% and 8.9%, respectively. A good correlation between the results of patient serum containing imipramine and desipramine simultaneously analyzed by ADx and gas liquid chromatography (GC) was observed, r2 = 0.964, n = 32. Results of patient serum containing amitriptyline and nortriptyline or doxepin and desmethyldoxepin analyzed by ADx and GC or GC-mass spectrometry were not well correlated; r2 = 0.738, n = 44 and r2 = 0.695, n = 21, respectively. The assay consistently flagged as "HI" serum with total imipramine and desipramine concentrations above 300 ng/mL by GC, and serum with greater than 360 ng/mL of amitriptyline and nortriptyline by GC/MS. No significant cross-reactivity was observed for drugs other than the TCA.
No takes yet. Share an insight, caveat, or question.
Poklis et al. (1990) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: