Key result
A dual optical photoplethysmography sensor using the Levenberg-Marquardt algorithm classified the degree of arteriovenous fistula stenosis with 94.82% accuracy and 92.22% precision.
Why the study?
Does a dual optical PPG sensor with Levenberg-Marquardt algorithm accurately classify the degree of AVF stenosis in hemodialysis patients?
Observational (n=11)
Does a dual optical PPG sensor with Levenberg-Marquardt algorithm accurately classify the degree of AVF stenosis in hemodialysis patients?
p-value: p=<0.05
A dual optical PPG sensor using the Levenberg-Marquardt algorithm provides high accuracy for noninvasive classification of AVF stenosis in hemodialysis patients.
May aid noninvasive AVF stenosis monitoring in HD; leaves open prospective validation before clinical adoption.
This paper proposes a noninvasive dual optical photoplethysmography (PPG) sensor to classify the degree of arteriovenous fistula (AVF) stenosis in hemodialysis (HD) patients. Dual PPG measurement node (DPMN) becomes the primary tool in this work for detecting abnormal narrowing vessel simultaneously in multi-beds monitoring patients. The mean and variance of Rising Slope (RS) and Falling Slope (FS) values between before and after HD treatment was used as the major features to classify AVF stenosis. Multilayer perceptron neural networks (MLPN) training algorithms are implemented for this analysis, which are the Levenberg-Marquardt, Scaled Conjugate Gradient, and Resilient Back-propagation, to identify the degree of HD patient stenosis. Eleven patients were recruited with mean age of 77 ± 10.8 years for analysis. The experimental results indicated that the variance of RS in the HD hand between before and after treatment was significant difference statistically to stenosis (p < 0.05). Levenberg-Marquardt algorithm (LMA) was significantly outperforms the other training algorithm. The classification accuracy and precision reached 94.82% and 92.22% respectively, thus this technique has a potential contribution to the early identification of stenosis for a medical diagnostic support system.
No takes yet. Share an insight, caveat, or question.
Du et al. (2018) conducted an observational in Arteriovenous fistula stenosis in hemodialysis patients (n=11). Dual optical photoplethysmography (PPG) sensor with Levenberg-Marquardt algorithm vs. Scaled Conjugate Gradient and Resilient Back-propagation algorithms was evaluated on Classification accuracy and precision of arteriovenous fistula stenosis (p=<0.05). A dual optical photoplethysmography sensor using the Levenberg-Marquardt algorithm classified the degree of arteriovenous fistula stenosis with 94.82% accuracy and 92.22% precision.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: