Microcantilevers modified with a self-assembled monolayer respond sensitively to specific ion concentrations. Here, we report the detection of trace amounts of CrO4(2-) using microcantilevers modified with a self-assembled monolayer of triethyl-12-mercaptododecylammonium bromide. The self-assembled monolayer was prepared on a silicon microcantilever coated with a thin layer of gold on one side. The microcantilever undergoes bending due to sorption of CrO4(2-) ions on the monolayer-modified side. It was found that a concentration of 10(-9) M CrO4(2-) can be detected using this technology in a flow cell. Other anions, such as Cl-, Br-, CO3(2-) (or HCO3-), and SO4(2-), have minimal effect on the deflection of this cantilever. The mechanics of the bending and the chemistry of cantilever modification are discussed.
No takes yet. Share an insight, caveat, or question.
Ji et al. (2001) studied this question.
Synapse has enriched 4 closely related papers on similar clinical questions. Consider them for comparative context: