Key result
Ultrasound strain imaging provided significantly higher image contrast between normal brain tissue and glial tumour tissue compared to conventional B-mode imaging (0.60 vs 0.39, p<0.0001).
Why the study?
Does ultrasound strain imaging improve image contrast between normal brain tissue and glial tumours compared to conventional B-mode imaging?
Observational (n=15)
No
Does ultrasound strain imaging improve image contrast between normal brain tissue and glial tumours compared to conventional B-mode imaging?
Answers are grounded in this paper and cite their sources.
Intraoperative ultrasound strain imaging provides significantly higher contrast between normal brain tissue and glial tumour tissue compared to conventional B-mode imaging.
BACKGROUND: Image contrast between normal tissue and brain tumours may sometimes appear to be low in intraoperative ultrasound. Ultrasound imaging of strain is an image modality that has been recently explored for intraoperative imaging of the brain. This study aims to investigate differences in image contrast between ultrasound brightness mode (B-mode) images and ultrasound strain magnitude images of brain tumours. METHODS: Ultrasound radiofrequency (RF) data was acquired during surgery in 15 patients with glial tumours. The data were subsequently processed to provide strain magnitude images. The contrast in the B-mode images and the strain images was determined in assumed normal brain tissue and tumour tissue at selected regions of interest (ROI). Three measurements of contrast were done in the ultrasound data for each patient. The B-mode and strain contrasts measurements were compared using the paired samples t- test. RESULTS: The statistical analysis of a total of 45 measurements shows that the contrasts in the strain magnitude images are significantly higher than in the conventional ultrasound B-mode images (P < 0.0001). CONCLUSIONS: The results indicate that ultrasound strain imaging provides better discrimination between normal brain tissue and glial tumour tissue than conventional ultrasound B-mode imaging. Ultrasound imaging of tissue strain therefore holds the potential of becoming a valuable adjunct to conventional intraoperative ultrasound imaging in brain tumour surgery.
No takes yet. Share an insight, caveat, or question.
Selbekk et al. (2012) conducted an observational in Glial brain tumours (n=15). Ultrasound strain imaging vs. Conventional ultrasound B-mode imaging was evaluated on Image contrast between normal brain tissue and tumour tissue (p=<0.0001). Ultrasound strain imaging provided significantly higher image contrast between normal brain tissue and glial tumour tissue compared to conventional B-mode imaging (0.60 vs 0.39, p<0.0001).