Answers are grounded in this paper and cite their sources.
PURPOSE: To describe recent technological additions to the NIDEK CXIII and Quest excimer lasers. METHODS: A summary article with data from previous published studies outlining the benefits of newer technology. RESULTS: The addition of a 1-kHz infrared eye tracker decreased the spread of laser spot placement from a mean of 228.79 microm without a tracker to 38.47 microm with the eye tracker. The addition of real-time torsion error correction produced a statistically significantly lower cylinder dispersion, mean manifest refractive cylinder, and error of angle postoperatively in eyes that underwent LASIK. CONCLUSIONS: The incorporation of an ultrahigh speed eye tracker and active cyclotorsion correction surpasses the minimal technology criteria required for accurate wavefront-based ablations.
No takes yet. Share an insight, caveat, or question.
George O. Waring (2009) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: