Answers are grounded in this paper and cite their sources.
In Response: The reports of tracheal tube injuries such as those by Hsu and Frogel (1,2), associated with Video laryngoscopes and specifically the GlideScope have focused attention on these devices and the skills required to use them. The path that we have followed to develop the GlideScope taught early that a definite learning curve exists to bring skill levels up to the optimal individual level. Professor Ron Walls, Women’s and Brigham Chair, Department of Emergency Medicine, proposes and teaches the following method of use for the GlideScope system: Step 1: First look at the mouth directly when introducing the GlideScope into the midline of the oral cavity. Step 2: Look at the monitor and elevate the tip of the blade to see the epiglottis and the glottic opening. Step 3: Look at the mouth to carefully guide the tube and stylet into position near the tip of the GlideScope blade. Step 4: Look at the monitor to complete the intubation under direct vision. I believe that this deliberate approach provides the method required to avoid injuries caused by Endotracheal tube advancement. In addition it is recommended by Verathon (GlideScope source) that a rigid, highly controllable Verathon stylet be used in conjunction with the GlideRite® style Endotracheal tube because the ski-shaped soft tip offers less potential to injure the pharyngeal or tracheal tissues. John Allen Pacey, MD, FRCS Verathon Medical (Canada) ULC 4224 Manor Street Burnaby, BC V5G 1B2 Vancouver, Canada
No takes yet. Share an insight, caveat, or question.
John Allen Pacey (2007) studied this question.