Answers are grounded in this paper and cite their sources.
In this study, we proposed a silicon nanobelt field-effect transistor (FET) to detect prostate specific antigen (PSA). The nanobelt FET device displayed n-channel depletion characteristics. The immobilization of prostate specific antibody (anti-PSA) molecules was attached onto the nanobelt FET surface by using the aldehyde groups of glutaraldehyde linked to the amino groups of 3-aminopropyltriethoxysilane (APTES). The shift in the drain current vs time curves of a nanobelt FET biosensor revealed that the electrical signal had a logarithmic relationship with respect to the concentration of the PSA, and detection capability was estimated in the 5 pg/mL level. To enhance the sensitivity of a nanobelt FET biosensor, this biosensor was designed by inserting arginine molecules between glutaraldehyde and APTES. Therefore, the detection capability of the developed sensor was extended to 50 fg/mL. Also, the relationship between the current shift and the logarithm of PSA concentration was exhibited linearity in the range 50 fg/mL - 500 pg/mL. The excellent electrical results of this label-free PSA nanobelt FET biosensor suggested that such biosensor might be potentially useful tools for biological research and future prostate cancer screening.
No takes yet. Share an insight, caveat, or question.
Wu et al. (2012) studied this question.