Key result
A simple microprocessor algorithm for ECG R-wave detection achieved a total error rate of approximately 1% for false positives and negatives when assessed on patient tapes.
Answers are grounded in this paper and cite their sources.
A simple microprocessor algorithm using gradient and duration criteria can reliably detect ECG R-waves with an error rate of approximately 1%.
ECG monitoring devices for coronary care units require reliable R-wave dectors as a prerequisite for subsequent analysis. A simple algorithm has been developed for the detection of R-waves in an ECG signal with a microprocessor (Motorola MC6800). Detection criteria used were gradient and duration of a single upslope or downslope. Rejection of muscle noise and other artifiacts was achieved by setting system constants to optimum values which were determined by a short preliminary trial. This constant threshold system has been shown to cope with sudden changes of QRS complex amplitude better than some self adjusting systems. Assessment runs using patient tapes from a coronary care unit show total errors due to false positives and negatives are typically about 1% of the actual number of R-wave occurrences.
No takes yet. Share an insight, caveat, or question.
Coleman et al. (1979) studied ECG monitoring in coronary care units. Microprocessor algorithm for R-wave detection vs. Self-adjusting systems was evaluated on Total errors due to false positives and negatives. A simple microprocessor algorithm for ECG R-wave detection achieved a total error rate of approximately 1% for false positives and negatives when assessed on patient tapes.