Key result
Medtronic and Guidant delivery systems for left ventricular pacing showed no significant differences in complication rates, including coronary sinus dissection (5% vs 6%, P=NS).
Why the study?
Do different coronary sinus leads and delivery systems have different complication rates in patients undergoing left ventricular pacing?
Observational (n=249)
Do different coronary sinus leads and delivery systems have different complication rates in patients undergoing left ventricular pacing?
Absolute Event Rate: 5% vs 6%
p-value: p=NS
There are no significant differences in complication rates between Medtronic and Guidant delivery systems and leads for left ventricular pacing.
Comparable complication rates between systems support similar safety profiles; observational data leaves open need for randomized confirmation before guiding selection.
AIMS: To compare different coronary sinus (CS) leads and delivery systems (DSs) for left ventricular pacing. METHODS AND RESULTS: Delivery systems-related (including CS dissection and dislocations during sheath/stylet removal) and lead-related (including failure to accomplish implantations and long-term malfunctions resulting in abandonment or repositioning/replacing of the lead) complications between systems and leads were compared. We used Medtronic (MDT) attain DS (n = 123) with over-the-wire (OTW) (4193, 4194) and stylet-driven (2187) leads, and Guidant (GDT) DS (n = 126) with Easytrak OTW leads (4513, 4518, and 4525). Coronory sinus dissection occurred in 6/123 (5%) cases using the MDT DS vs. 7/126 (6%) with GDT DS (P= NS). Dislocations during sheath/stylet removal occurred in 8/123 cases (6%) with MDT DS, and in 8/126 (6%) with GDT DS (P= NS). Failure to achieve successful implantation occurred in 6/32 (19%) of the 2187 leads, in 11/87(13%) of the 4193/4194 leads, in 7/94(7%) of the 4513/4518 leads, and in 4/29 (14%) of the 4525 leads (P= NS). Long-term lead-related complications occurred in 5/32 (15%) of the 2187 leads, 19/80 (23%) of the 4193/4194 leads, 19/93 (20%) of the 4513/4518 leads, and 2/28 (7%) of the 4525 leads (P= NS). CONCLUSION: No significant differences in complication rates between systems and leads were observed.
No takes yet. Share an insight, caveat, or question.
Nof et al. (2007) conducted an observational in Left ventricular pacing (n=249). Medtronic (MDT) attain delivery system and leads vs. Guidant (GDT) delivery system and leads was evaluated on Coronary sinus dissection (p=NS). Medtronic and Guidant delivery systems for left ventricular pacing showed no significant differences in complication rates, including coronary sinus dissection (5% vs 6%, P=NS).