Key result
The OmniaSecure defibrillation lead achieved 97.5% successful defibrillation at implantation and 97.1% freedom from lead-related major complications at 6 and 12 months.
Why the study?
With implantable cardioverter-defibrillators lasting longer, interest is growing in reliable leads with targeted placement, prompting evaluation of the novel small-diameter OmniaSecure lead.
Does the OmniaSecure defibrillation lead provide safe and effective defibrillation in patients indicated for de novo ICD or CRT-D implantation?
Does the OmniaSecure defibrillation lead provide safe and effective defibrillation in patients indicated for de novo ICD or CRT-D implantation?
The novel small-diameter OmniaSecure defibrillation lead demonstrated high defibrillation success and a low rate of lead-related major complications with zero fractures at 12 months.
Supports OmniaSecure lead consideration in de novo ICD/CRT-D; leaves open randomized confirmation of long-term outcomes.
BACKGROUND: Implantable cardioverter-defibrillators last longer, and interest in reliable leads with targeted lead placement is growing. The OmniaSecure defibrillation lead is a novel, small-diameter, catheter-delivered lead designed for targeted placement, based on the established SelectSecure SureScan MRI Model 3830 lumenless pacing lead platform. OBJECTIVE: This trial assessed safety and efficacy of the OmniaSecure defibrillation lead. METHODS: The worldwide LEADR pivotal clinical trial enrolled patients indicated for de novo implantation of a primary or secondary prevention implantable cardioverter-defibrillator or cardiac resynchronization therapy defibrillator, all of whom received the study lead. The primary efficacy end point was successful defibrillation at implantation per protocol. The primary safety end point was freedom from study lead-related major complications at 6 months. The primary efficacy and safety objectives were met if the lower bound of the 2-sided 95% credible interval was >88% and >90%, respectively. RESULTS: In total, 643 patients successfully received the study lead, and 505 patients have completed 12-month follow-up. The lead was placed in the desired right ventricular location in 99.5% of patients. Defibrillation testing at implantation was completed in 119 patients, with success in 97.5%. The Kaplan-Meier estimated freedom from study lead-related major complications was 97.1% at 6 and 12 months. The trial exceeded the primary efficacy and safety objective thresholds. There were zero study lead fractures and electrical performance was stable throughout the mean follow-up of 12.7 ± 4.8 months (mean ± SD). CONCLUSION: The OmniaSecure lead exceeded prespecified primary end point performance goals for safety and efficacy, demonstrating high defibrillation success and a low occurrence of lead-related major complications with zero lead fractures.
No takes yet. Share an insight, caveat, or question.
Crossley et al. (2024) studied Indication for de novo implantation of a primary or secondary prevention ICD or CRT-D (n=643). OmniaSecure defibrillation lead was evaluated on Successful defibrillation at implantation per protocol. The OmniaSecure defibrillation lead achieved 97.5% successful defibrillation at implantation and 97.1% freedom from lead-related major complications at 6 and 12 months.