Key result
In an 80-year-old woman with a VDD-pacemaker, AV-synchrony of 58-73% was maintained despite displacement of the atrial dipole into the right ventricular outflow tract.
Case Report (n=1)
This case report highlights that AV-synchrony can be partially maintained by left atrial sensing when a VDD pacemaker's atrial dipole is displaced into the right ventricular outflow tract.
Case of preserved AV synchrony with malpositioned VDD dipole is hypothesis-generating; does not support altering lead management.
The case of an 80-year-old woman who underwent permanent VDD-pacemaker implantation for recurrent syncope in the presence of second-degree type 2 AV-block is reported. During follow-up, low atrial sensing with AV-synchrony of only 58-73% was noted. Four years after the pacemaker implantation, the patient was hospitalized for non-cardiac reasons and the chest radiograph showed displacement of the atrial dipole into the right ventricular outflow tract (RVOT). It is hypothesized that AV-synchrony was maintained by left atrial sensing due to the anatomic proximity of the RVOT to the left atrial appendage.
No takes yet. Share an insight, caveat, or question.
Schumann et al. (2007) conducted a case report in Recurrent syncope with second-degree type 2 AV-block (n=1). VDD-pacemaker implantation was evaluated on AV-synchrony. In an 80-year-old woman with a VDD-pacemaker, AV-synchrony of 58-73% was maintained despite displacement of the atrial dipole into the right ventricular outflow tract.