Key result
Increased EtCO2 (40-45 mmHg) significantly reduced the trending ability of SpHb during laparoscopic surgery, showing a concordance rate of 77.8% vs 95.9% in the low EtCO2 group.
Why the study?
The study evaluated the impact of increased end-tidal carbon dioxide on the accuracy and trending ability of continuous noninvasive hemoglobin monitoring in laparoscopic surgery.
Does increased end-tidal carbon dioxide reduce the accuracy and trending ability of continuous noninvasive hemoglobin monitoring in patients undergoing laparoscopic surgery?
RCT (n=64)
randomly allocated
Does increased end-tidal carbon dioxide reduce the accuracy and trending ability of continuous noninvasive hemoglobin monitoring in patients undergoing laparoscopic surgery?
Absolute Event Rate: 77.8% vs 95.9%
Increased end-tidal carbon dioxide significantly reduces the trending ability of continuous noninvasive hemoglobin monitoring during laparoscopic surgery.
The pulse CO-Oximetry allows continuous, noninvasive monitoring of hemoglobin (SpHb). We assessed the impact of increased end-tidal carbon dioxide (EtCO2) on the accuracy and trending ability of SpHb in laparoscopic surgery. Participants (n = 64) were randomly allocated to the low carbon dioxide (CO2) group (EtCO2: 30–35 mmHg) or the high CO2 group (EtCO2: 40–45 mmHg). The SpHb and laboratory hemoglobin (tHb) were obtained during surgery. The correlation coefficient (r) between SpHb and tHb showed greater tendency in the low CO2 group (r = 0.68) than in the high CO2 group (r = 0.43). The bias (precision) was −1.18 (1.09) with a limit of agreement (LOA) of −3.31 to 0.95 in low CO2 group and −1.02 (1.24) with a LOA of −3.45 to 1.42 in high CO2 group; they did not differ significantly between the groups (p = 0.246). The low CO2 group showed a high concordance rate of 95.9% and a moderate correlation between ΔSpHb and ΔtHb (r = 0.53). However, the high CO2 group showed a concordance rate of 77.8% and no correlation between ΔSpHb and ΔtHb (r = 0.11). In conclusion, increased EtCO2 significantly reduced the trending ability of SpHb during laparoscopic surgery. Caution should be executed when interpreting SpHb values during laparoscopic surgery in patients with hypercapnia.
No takes yet. Share an insight, caveat, or question.
Kim et al. (2022) conducted an RCT in laparoscopic gastrectomy (n=64). High end-tidal carbon dioxide (EtCO2) vs. Low end-tidal carbon dioxide (EtCO2: 30-35 mmHg) was evaluated on concordance rate between ΔSpHb and ΔtHb. Increased EtCO2 (40-45 mmHg) significantly reduced the trending ability of SpHb during laparoscopic surgery, showing a concordance rate of 77.8% vs 95.9% in the low EtCO2 group.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: