Answers are grounded in this paper and cite their sources.
Fast- and slow-neutron contamination in an 18-MV photon beam from a Philips SL/75-20 linear accelerator has been measured. Aluminum and indium foils were activated to determine fast- and slow-neutron fluence, which were largely independent of field sizes. Measured fast-neutron fluences were typically 13.9 X 10(4) and 4.4 X 10(4) neutrons/cm2/rad of x ray inside and 5 cm outside the field, respectively. Slow-neutron fluences, 1.3 X 10(4) neutrons/cm2/rad of x ray, remained relatively constant inside and outside the field. The reported results are about three times higher than neutron fluences recently reported with a betatron operated at the same energy.
No takes yet. Share an insight, caveat, or question.
Gur et al. (1978) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: