This paper demonstrates a calorimetric procedure for calibrating high-heat-flux sensors. The results are in agreement with calibrations obtained using black-body radiation. However, the proposed method has the potential of being more accurate than traditional approaches. This new procedure calibrates sensors to measure correctly under conditions of concentrated solar radiation. At present, the thermal balance calibration technique in the laboratory is limited to solar irradiances of approximately 100 kW m −2 . The next step is to demonstrate this methodology to higher irradiances under non-laboratory conditions in the CIEMAT solar furnace at Plataforma Solar de Almería.