The diastereoisomeric separation of peptidomimetics of hexarelin, a strong growth hormone secretagogue, in CE has been studied. Highly sulfated-gamma-CD was found to be an appropriate selector for the separation of the stereoisomers. However, non-repeatable analyses were obtained on bare fused silica capillary due to the progressive adsorption of the analytes on the capillary wall. Two types of polyelectrolyte coating agents were tested to prevent this phenomenon. Coating with neutral polyethylene oxide was found to be efficient but resulted in a very long analysis time (about 40 min). Coating with cationic poly(diallyldimethylammonium) chloride was found both to prevent analyte adsorption, reduce analysis time and alter separation selectivity. EOF measurement revealed that the highly sulfated-gamma-CDs were strongly adsorbed on the poly(diallyldimethylammonium) chloride coating surface yielding a stable strong cathodic EOF, which considerably reduced analysis time (about 12 min). Very good repeatability of analysis was obtained (RSD(migration time)<1%).