Describes an active RC filter design to eliminate 60 cps power-frequency interference in biological potential amplifiers like EKGs.
Answers are grounded in this paper and cite their sources.
May attenuate 60 cps interference in ECG amplifiers; leaves open validation in modern digital systems.
An active RC filter with an m-derived low-pass response is described. With a cutoff of 45 cps and a sharp attenuation peak at 60 cps, it is particularly suitable for eliminating power-frequency interference from EKG, EEG, or other biological potential amplifiers. It is small in size and requires only one 12AX7 or its transistor equivalent for the active elements. Typical design data and performance curves are given.
No takes yet. Share an insight, caveat, or question.
Gifford White (1961) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: