Answers are grounded in this paper and cite their sources.
A drift method is described for measuring specific heats in intense magnetic fields at low temperatures. Capacitance thermometry is used, and an automated data-collection system utilizes the imbalance of a transformer-ratio-arm bridge to process the capacitance data. The zero-field specific heat must be known, and measurement of the in situ drift in zero-field calibrates the thermal link. Additional calibration of the link is required if the link’s magnetothermal conductivity effects are significant, and a specific example of a copper-wire link is presented. The method resolves complex structure in the specific heat near a steep λ-type anomaly, as illustrated by measurements on a chromite spinel at 7.5 T. The uncertainty in the method is estimated to be ≂±7%.
No takes yet. Share an insight, caveat, or question.
Lawless et al. (1982) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: