Answers are grounded in this paper and cite their sources.
An oxidase-based glucose sensor has been developed that uses a mercaptosilane-modified platinum electrode to achieve selectivity of electrochemical interferants. A platinum-iridium (9:1) wire (0.178 mm o.d., sensing area of 1.12 mm2) is modified with (3-mercaptopropyl)trimethoxysilane. The modified sensors show excellent operational stability for more than 5 days. Glucose oxidase is immobilized on the modified surface (i) by using 3-maleimidopropionic acid as a linker or (ii) by cross-liking with bovine serum albumin using glutaraldehyde. Sensitivities in the range of 9.97 nA/mM glucose are observed when the enzyme is immobilized by method ii. Lower sensitivities (1.13 x 10(-1) nA/mM glucose) are observed when immobilization method i is employed. In terms of linear response range, the sensor enzyme-immobilized by method i is superior to that immobilized by method ii. The linearity is improved upon coating the enzyme layer with polyurethane. The sensor immobilized by method ii and coated with polyurethane exhibits a linear range to 15 mM glucose and excellent selectivity to glucose (0.47 nA/mM) against interferants such as ascorbic acid, uric acid, and acetaminophen.
No takes yet. Share an insight, caveat, or question.
Jung et al. (1996) studied this question.