The positron-emitting activation of tissue by α-particle beams has been studied, using the annihilation γ-ray coincidence detection technique, and the Anger positron camera. The decay spectrum in tissue following irradiation by 53 Mev α-particles from the Berkeley 88-inch cyclotron has been measured, as well as a visualization of the activation pattern resulting from a 910 Mev α-particle beam stopping in a tissue-like plastic phantom, at the Berkeley 184-inch synchrocyclotron. Also studied were the activation patterns resulting from transmission of 910 MeV α-beams through a wax head phantom, and the activation pattern in patients' heads resulting from one sitting for the pituitary ablation therapy technique. With slight improvements in resolution and sensitivity, an `on-line' positron camera technique could be useful for alignment of the end of the beam track in Bragg- peak therapy with heavy charged particles. There are other novel applications which might be possible, using tissue activation techniques.
No takes yet. Share an insight, caveat, or question.
Maccabee et al. (1969) studied this question.