Key result
Large-eddy simulation of an axial ventricular assist device identified universal local flow regions and structures with significant turbulent stresses that could lead to blood damage.
Why the study?
Turbulence occurring in VADs and local turbulent structures causing increased turbulent stresses had not yet been analyzed in detail to improve hemocompatibility.
This computational study identifies specific regions of high turbulent stress in axial ventricular assist devices, providing a basis for design optimizations to improve hemocompatibility and reduce blood damage.
Answers are grounded in this paper and cite their sources.
Identified turbulent regions may guide axial VAD design; leaves open clinical translation from Level 5 computational data.
Numerical flow simulations that analyze the turbulent flow characteristics within a turbopump are important for optimizing the efficiency of such machines. In the case of ventricular assist devices (VADs), turbulent flow characteristics must be also examined in order to improve hemocompatibility. Turbulence increases the shear stresses in the VAD flow, which can lead to an increased damage to the transported blood components. Therefore, an understanding of the turbulent flow patterns and their significance for the numerical blood damage prediction is particularly important for flow optimizations in VADs in order to identify and thus minimize flow regions where blood could be damaged due to high turbulent stresses. Nevertheless, the turbulence occurring in VADs and the local turbulent structures that lead to increased turbulent stresses have not yet been analyzed in detail in these machines. Therefore, this study aims to investigate the turbulence in an axial VAD in a comprehensive and double tracked way. First, the flow in an axial VAD was computed using the large-eddy simulation method, and it was verified that the majority of the turbulence was directly resolved by the simulation. Then, the turbulent flow state of the VAD was quantified globally. For this purpose, a self-designed evaluation method, the power loss analysis, was used. Subsequently, local flow regions and flow structures were identified where significant turbulent stresses prevail. It will be shown that the identified regions are universal and will also occur in other axial blood pumps as well, for example, in the HeartMate II.
No takes yet. Share an insight, caveat, or question.
Torner et al. (2020) studied this question. Large-eddy simulation was evaluated on Turbulent flow characteristics and power loss. Large-eddy simulation of an axial ventricular assist device identified universal local flow regions and structures with significant turbulent stresses that could lead to blood damage.