Key result
A novel quality index for PPG-based heart rate signals showed high accuracy for estimating mean HR but poor to moderate accuracy for HRV features compared to ECG.
Why the study?
Real-life wearable heart rate measurements suffer from motion artifacts and frequent data loss, particularly in photoplethysmography devices.
Does a real-time quality index improve the accuracy of heart rate variability feature extraction in PPG-based wearable cardiac sensors?
Observational
Does a real-time quality index improve the accuracy of heart rate variability feature extraction in PPG-based wearable cardiac sensors?
A novel quality index can identify reliable time windows in PPG-based cardiac monitoring, reducing errors in heart rate variability feature extraction caused by motion artifacts.
Answers are grounded in this paper and cite their sources.
May support mean HR monitoring from PPG wearables; leaves open reliable HRV extraction pending validation.
Wearable cardiac sensors pave the way to advanced cardiac monitoring applications based on heart rate variability (HRV). In real-life settings, heart rate (HR) measurements are subject to mo-tion artifacts that can be timely removed from the recordings. This leads to frequent data loss in the HR signal, especially for commercial devices based on photoplethysmography (PPG). The cur-rent study had two main goals: (i) to provide a white-box quality index that estimates the amount of missing samples in any piece of HR signal; and (ii) to quantify the impact of data loss on feature extraction in a PPG-based HR signal. This was done by comparing real-life recordings from com-mercial sensors featuring both PPG (Empatica E4) and ECG (Zephyr BioHarness 3). After an out-lier rejection process, our quality index was used to isolate portions of ECG-based HR signal that could be used as benchmark, to validate the output of Empatica E4 at the signal level and at the feature level. Our results showed high accuracy for estimating the mean HR, poor accuracy for short-term HRV features and moderate accuracy for longer-term HRV features. Levels of error could be substantially reduced by using our quality index to identify time windows with few or no missing data.
No takes yet. Share an insight, caveat, or question.
Vila et al. (2021) conducted an observational in Cardiac monitoring. PPG-based HR signal (Empatica E4) vs. ECG-based HR signal (Zephyr BioHarness 3) was evaluated on Accuracy for estimating mean HR and HRV features. A novel quality index for PPG-based heart rate signals showed high accuracy for estimating mean HR but poor to moderate accuracy for HRV features compared to ECG.