Key result
Strain-gauged hip replacement with FM transmitter successfully records in vivo axial loads over 40 days.
Why the study?
Can in vivo hip joint forces be recorded using a strain gauged 'English' prosthesis with an implanted transmitter?
Can in vivo hip joint forces be recorded using a strain gauged 'English' prosthesis with an implanted transmitter?
The study demonstrates the feasibility of measuring in vivo hip joint forces using an instrumented prosthesis with an implanted FM transmitter.
Demonstrates single-case feasibility of implanted telemetry for hip loads; leaves open prospective validation before clinical integration.
This paper describes the results obtained from implanting a strain gauged version of an ‘English’ hip joint replacement together with a totally implantable FM radio transmitter. The implant is based upon a new concept in the design of femoral hip components having a diminished head offset to reduce head load and improved stem shape permitting alignment of the neck along the theoretical axis of peak load transmitted during the gait cycle. The implant was inserted using the ‘English’ trochanteric approach (English, 1975) which further reduces the load on a prosthetic hip joint with the use of a spacer out from the redundant femoral head to rearrange the trochanteric muscle lever arms. The resulting axial load is detected by four strain gauges mounted on a ‘piston in cylinder’ arrangement contained within the thickened neck of the prosthesis. The single channel FM transmitter relays the gauge output to a signal processing unit to give a direct output of activity for recording on a UV recorder. Recordings were taken during implantation, recovery, walking (at three days) physiotherapy, stair climbing and walking over a period of forty days.
No takes yet. Share an insight, caveat, or question.
Kilvington et al. (1981) studied Hip joint replacement. Strain gauged 'English' hip joint replacement with FM radio transmitter was evaluated on In vivo hip joint forces (axial load). A strain gauged 'English' hip joint replacement with an implanted FM radio transmitter successfully recorded in vivo axial loads during various activities over a 40-day period.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: