Answers are grounded in this paper and cite their sources.
An on-line flow injection method for the direct determination of silver in sea-water using isotope dilution inductively coupled plasma mass spectrometry (ID-ICP-MS) is described. A mini-column packed with 50 µl Dowex 1-X8 anion exchange resin was used to separate and concentrate Ag from sea-water following on-line merging with 0.5 M HCl. Recoveries of 100±1% and 96±1% were obtained for NRC CRMs SLEW-3 and CASS-4, respectively. One run can be completed in less than 6 min using a 12 ml volume of sea-water. The detection limit (3σ), estimated at 0.06 pg ml−1, is superior to previously published methods. Dissolved Ag concentrations of 1.93±0.10, 5.15±0.10 and 5.42±0.10 pg ml−1 were obtained in SLEW-3, CASS-2 and CASS-4 CRMs, respectively. The method was also assessed for determination of Ag in a marine sediment CRM MESS-3, and a concentration of 0.169±0.002 µg g−1 was obtained using ID ICP-MS after an off-line column separation, in agreement with the certified value of 0.18±0.02 µg g−1. The on-line method was also applied to analyses of sea-water and sufficiently diluted digests of marine sediments using an external calibration technique. Concentrations of 1.96±0.10 pg ml−1 (mean and one standard deviation, n = 3) and 5.55±0.15 (n = 3) pg ml−1 were obtained for dissolved Ag in SLEW-3 and CASS-4, respectively, in agreement with results generated by ID ICP-MS. A concentration of 0.181±0.004 (n = 3) µg g−1 in MESS-3 was obtained, based on 200-fold diluted digests, in agreement with the certified value.
No takes yet. Share an insight, caveat, or question.
Yang et al. (2002) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: