Key result
Echocardiographic measures of left atrial strain and morphology significantly correlated with the atrial contraction signal (A4 amplitude) in 21 patients with Micra-AV leadless pacemakers.
Why the study?
To identify which patients leadless pacemakers can guarantee atrio-ventricular synchrony in, by exploring correlations between echocardiographic LA size and function with A4 amplitude in Micra-AV patients.
Does echocardiographic assessment of left atrial strain predict adequate A4 sensing and atrio-ventricular synchrony in patients implanted with a Micra-AV leadless pacemaker?
Observational (n=21)
Does echocardiographic assessment of left atrial strain predict adequate A4 sensing and atrio-ventricular synchrony in patients implanted with a Micra-AV leadless pacemaker?
Preimplant echocardiographic evaluation of left atrial strain can predict adequate A4 sensing and atrio-ventricular synchrony in patients receiving leadless pacemakers.
INTRODUCTION: The importance of atrio-ventricular synchrony pacing in sinus rhythm patients is known. To identify patients in whom leadless pacemakers are able to guarantee this atrio-ventricular synchrony, we explored correlations among echocardiographic measures of left atrial (LA) size and function (doppler parameter and strain) with A4 amplitude in patients implanted with new generation Micra-AV device. METHODS: After implantation with Micra-AV system, patients underwent device interrogation to evaluate AV synchrony based on the sensing of atrial mechanics and echocardiographic exam to assess LA morphology and LA function. RESULTS: In the 21 studied patients (14 males, 72 ± 13 years), the A4 wave amplitude values inversely correlated with LA antero-posterior diameter, LA volume, LA contraction strain and LA conduit strain, while they were positively related with LA reservoir strain. DISCUSSION: Our results indicate a statistically significant relationship between morphological echocardiographic LA parameters and atrial contraction signal (A4), detected by leadless pacemakers and used to synchronize ventricular pacing with the atrium. Instantaneous LA function assessment obtained with LA strain provides incremental information over morphological parameters. LA strain evaluates atrial myocardial deformation during the whole cardiac cycle. We found higher value of A4 in patients that have grater absolute value of LAsr, LAscd and LAsct, that are simple and measurable parameters of LA functional capacity. CONCLUSION: Preimplant echocardiographic evaluation of the atrial contractility may be useful in predicting adequate A4 sensing and consequently a good atrio-ventricular synchrony pacing. Echocardiography LA strain study seems promising in Micra-AV patient selection.
No takes yet. Share an insight, caveat, or question.
Troisi et al. (2023) conducted an observational in Sinus rhythm patients implanted with Micra-AV device (n=21). Left atrial strain and morphological parameters was evaluated on Correlation between echocardiographic measures of left atrial size/function and A4 amplitude. Echocardiographic measures of left atrial strain and morphology significantly correlated with the atrial contraction signal (A4 amplitude) in 21 patients with Micra-AV leadless pacemakers.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: