The 23 mm ALLEGRA valve maintained favorable gradients (<20 mmHg) in surgical valves ≥21 mm regardless of implant depth, though leaflet pinwheeling worsened at lower implantation depths.
Why the study?
The impact of implant depth on hydrodynamic function following valve-in-valve intervention with the ALLEGRA transcatheter heart valve across different surgical valve designs was unknown.
Does implantation depth affect the hydrodynamic function of the ALLEGRA THV during valve-in-valve interventions in various surgical bioprostheses?
Does implantation depth affect the hydrodynamic function of the ALLEGRA THV during valve-in-valve interventions in various surgical bioprostheses?
Answers are grounded in this paper and cite their sources.
In an in vitro valve-in-valve model, the 23 mm ALLEGRA THV demonstrated favorable hemodynamics in surgical valves ≥21 mm regardless of implant depth, but lower implantation depths worsened leaflet pinwheeling.
AIMS: We aimed to assess the impact of implant depth on hydrodynamic function following valve-in-valve (VIV) intervention using the ALLEGRA transcatheter heart valve (THV) in three different surgical valve designs. METHODS AND RESULTS: Multiple implantation depths (+2 mm, -2 mm and -6 mm) were tested using a 23 mm ALLEGRA THV for VIV intervention in 19 mm, 21 mm, 23 mm, and 25 mm Epic, Mitroflow and Magna Ease bioprosthetic valves. Multimodality imaging and hydrodynamic evaluation was performed at each implantation depth. The 23 mm ALLEGRA valve had gradients <20 mmHg in the Mitroflow and Epic valves sized ≥21 mm, and in all sizes of the Magna Ease valve. Gradients did not increase significantly at lower implantation depths. The 19 mm Epic (+2 mm: 20.1±0.6 mmHg, -2 mm: 18.8±0.5 mmHg, -6 mm: 22.8±0.3 mmHg) and 19 mm Mitroflow (+2 mm: 24.1±0.2 mmHg, -2 mm: 31.5±0.3 mmHg, -6 mm: 25.6±0.2 mmHg) valves had elevated mean gradients. In larger sized surgical valves (≥23 mm) the regurgitant fraction was higher at low implantation depths. Pinwheeling was significantly worse in the smaller sized (≤21 mm) surgical valves and also at low (<-2 mm) implantation depth. CONCLUSIONS: The 23 mm ALLEGRA valve had favourable (<20 mmHg) gradients in all surgical valves sized ≥21 mm, even when the THV was implanted low. In 19 mm sized Mitroflow and Epic valves, gradients were elevated (>20 mmHg). While there was no major difference in mean transvalvular gradients, leaflet pinwheeling was worse at lower implantation depths.
No takes yet. Share an insight, caveat, or question.
Sathananthan et al. (2020) studied Valve-in-valve intervention. ALLEGRA transcatheter heart valve implantation at different depths vs. Different implantation depths was evaluated on Hydrodynamic function (transvalvular gradients, regurgitant fraction, pinwheeling). The 23 mm ALLEGRA valve maintained favorable gradients (<20 mmHg) in surgical valves ≥21 mm regardless of implant depth, though leaflet pinwheeling worsened at lower implantation depths.