Key result
Supra-annular (high) positioning of the CoreValve THV in small surgical bioprostheses reduced residual gradients by 25% compared to low positioning, though outcomes were highly model-dependent.
Why the study?
Does the size and position of a self-expanding transcatheter heart valve affect hemodynamics in small surgical bioprostheses in an in vitro model?
Does the size and position of a self-expanding transcatheter heart valve affect hemodynamics in small surgical bioprostheses in an in vitro model?
In an in vitro model, supra-annular positioning of self-expanding transcatheter heart valves improves hemodynamics in small surgical bioprostheses, but risks like regurgitation and embolization vary by valve model.
Answers are grounded in this paper and cite their sources.
Supra-annular CoreValve positioning may optimize ViV hemodynamics in small BP; leaves open need for prospective clinical validation.
AIMS: The valve-in-valve (ViV) procedure has become a valuable alternative for the treatment of failed surgical bioprostheses (BP) in high-risk patients. However, in small BP, the clinical outcomes have been suboptimal due to high post-procedural gradients. We aimed to examine the effect of size and position of the self-expanding transcatheter heart valve (THV) CoreValve on the haemodynamics of ViV within small BP. METHODS AND RESULTS: Sizes 23 and 26 mm of the CoreValve were implanted in sizes 19 and 21 mm of three BP models: Trifecta, Mitroflow and Epic Supra. The THV was tested in three positions -normal (manufacturer recommendation), low (4 mm below normal) and high (4 mm above normal)- using a pulse duplicator. Haemodynamics were assessed by Doppler echocardiography and flowmeter, and GOA with a high-speed camera. Higher implantation was associated with lower residual gradients (normal position: -9%, high: -25% versus low). High position was, however, associated with increased risk of regurgitation in the Mitroflow and embolisation in the Epic Supra. Using a 26 mm THV instead of a 23 mm was associated with larger EOAs in the Trifecta, smaller in the Mitroflow, and increased risk of embolisation in the Epic Supra. CONCLUSIONS: Supra-annular positioning of the CoreValve THV is associated with improved post-ViV haemodynamics in small surgical BP. The haemodynamic outcomes are highly dependent on the model and size of surgical BP.
No takes yet. Share an insight, caveat, or question.
Zenses et al. (2018) studied Failed surgical bioprostheses. CoreValve transcatheter heart valve size and positioning vs. Different sizes (23 vs 26 mm) and positions (normal, low, high) was evaluated on Haemodynamics (residual gradients, regurgitation, embolisation, EOA). Supra-annular (high) positioning of the CoreValve THV in small surgical bioprostheses reduced residual gradients by 25% compared to low positioning, though outcomes were highly model-dependent.
Synapse has enriched 10 related papers. Consider them for comparative context: