This is the authors' abstract. We don't add key points for this paper.
When making linear or nonlinear conductivity measurements it is often desirable to apply a high magnetic field locally to a sample. If the whole magnet assembly is integral to the sample probe, then, particularly when making low temperature measurements, a low total mass is desired. We have designed, modeled, and tested a variable magnetic field sweep unit suitable for use over a wide temperature range from cryogenic to ≈200 °C which provides a sweepable uniform magnetic field over an air gap of 2 mm and spatial extent of 4 mm×4 mm. The unit consists of a magnet yoke structure with a mass of 18 g and a spur gear driven rotatable magnet to vary the gap field in a nearly sinusoidal manner as a function of the magnet rotation angle. In the present design, a 20 g SmCo magnet has been used which allows for low temperature operation to ≈10 K, when attached to a cryogenic refrigerator cold finger. The shape of the magnetic yoke structure has been modeled and optimized using three-dimensional magnetic field software. The gap field uniformity can thus be modeled and tested experimentally. In a present working model the field for a 2 mm gap at room temperature B(θ)=1.05⋅cos(θ)(T) where θ is the magnet rotation angle. The field sweep amplitude at 19 K only increased by 3% over the 295 K value. With 0.25-mm-thick permunder pole tips, the field amplitude in a similar unit was increased to 1.20 T over a gap of 1.5 mm.