Key result
Bayer Immuno 1 troponin assay achieves 100% sensitivity for MI twelve hours after admission.
Why the study?
Does the Bayer Immuno 1 immunoassay accurately measure cardiac troponin I and diagnose myocardial infarction in patients with acute chest pain?
Observational (n=190)
Does the Bayer Immuno 1 immunoassay accurately measure cardiac troponin I and diagnose myocardial infarction in patients with acute chest pain?
The Bayer Immuno 1 troponin I assay demonstrates acceptable analytical performance and achieves 100% sensitivity for myocardial infarction by 12 hours after admission.
Supports troponin I testing at 12 hours to rule out MI; leaves open assay comparability and interference effects in broader populations.
We have evaluated the analytical and clinical performance of an automated immunoassay for serum cardiac troponin I (Bayer Immuno 1TM, Bayer Diagnostics, Tarrytown, NY). The between batch imprecision was found to be between 1.2 and 3.2% over the concentration range 2.5 - 34.0 microg/L. The analytical range obtained from duplicate analysis of patient samples and defined as a coefficient of variation of 10% or less was 0.3 - 200 microg/L. The detection limit was found to be less than 0.1 microg/L. A method comparison with the Dade Stratus method (Dade Behring, Wilmington, DE) yielded regression statistics with a slope of 0.705 and an intercept of -0.260. An analysis of samples from 40 patients with renal failure demonstrated six with detectable levels of troponin I (0.2 - 1.9 microg/L). Samples from patients with paraproteinaemia did not demonstrate detectable troponin I (from n = 30); however, two patients with elevated rheumatoid factor titers (from n = 20) demonstrated a detectable amount of troponin I (0.1 and 0.2 microg/L). In a study of 100 patients admitted with acute chest pain and a diagnosis of unstable angina, 6 were subsequently diagnosed as having suffered a myocardial infarction. On admission the sensitivity and specificity of the troponin I results were 26.7% and 94.7%, respectively, moving to 100% and 83% 12 hours after admission.
No takes yet. Share an insight, caveat, or question.
Clark et al. (1999) conducted an observational in Acute chest pain, renal failure, paraproteinaemia, elevated rheumatoid factor (n=190). Bayer Immuno 1 automated immunoassay for serum cardiac troponin I vs. Dade Stratus method was evaluated on Sensitivity and specificity for myocardial infarction.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: