We have micromachined a mechanical sensor that uses interferometry to detect the differential and absolute deflections of two adjacent cantilevers. The overall geometry of the device allows simple fluidic delivery to each cantilever to immobilize molecules for biological and chemical detection. We show that differential sensing is 50 times less affected by ambient temperature changes than the absolute, thus enabling a more reliable differentiation between specific cantilever bending and background effects. We describe the fabrication process and show results related to the dynamic characterization of the device as a differential sensor. The root-mean-squared (r.m.s.) sensor noise in water and air is /spl sim/1 nm over the frequency range of 0.4-40 Hz. We also find that in air, the deflection resolution is limited only by the cantilever's thermomechanical noise level of 0.008 /spl Aring//Hz/sup 1/2/ over the frequency range of 40-1000 Hz.