Key result
Operating an IABP at a 45° angle slows deflation by ~35% vs. horizontal positioning.
Why the study?
Does operating an intra-aortic balloon pump at a 45° angle reduce balloon wall movement, pressure pulse generation, and volume displacement compared to a 0° horizontal position in an in vitro mock aorta?
Does operating an intra-aortic balloon pump at a 45° angle reduce balloon wall movement, pressure pulse generation, and volume displacement compared to a 0° horizontal position in an in vitro mock aorta?
In an in vitro mock circulation model, operating an intra-aortic balloon pump at a 45° semi-recumbent angle significantly delays balloon tip deflation and reduces deflation pressure pulse and suctioned volume compared to the horizontal position, potentially compromising ventricular afterload reduction.
Answers are grounded in this paper and cite their sources.
May impair IABP afterload reduction at 45°; leaves open clinical relevance of angulation in patients.
The intra-aortic balloon pump (IABP) is a ventricular assist device that is used with a broad range of pre-, intra-, and postoperative patients undergoing cardiac surgery. Although the clinical efficacy of the IABP is well documented, the question of reduced efficacy when patients are nursed in the semi-recumbent position remains outstanding. The aim of the present work is therefore to investigate the underlying mechanics responsible for the loss of IABP performance when operated at an angle to the horizontal. Simultaneous recordings of balloon wall movement, providing an estimate of its diameter (D), and fluid pressure were taken at three sites along the intra-aortic balloon (IAB) at 0 and 45°. Flow rate, used for the calculation of displaced volume, was also recorded distal to the tip of the balloon. An in vitro experimental setup was used, featuring physiological impedances on either side of the IAB ends. IAB inflation at an angle of 45° showed that D increases at the tip of the IAB first, presenting a resistance to the flow displaced away from the tip of the balloon. The duration of inflation decreased by 15.5%, the inflation pressure pulse decreased by 9.6%, and volume decreased by 2.5%. Similarly, changing the position of the balloon from 0 to 45°, the balloon deflation became slower by 35%, deflation pressure pulse decreased by 14.7%, and volume suctioned was decreased by 15.2%. IAB wall movement showed that operating at 45° results in slower deflation compared with 0°. Slow wall movement, and changes in inflation and deflation onsets, result in a decreased volume displacement and pressure pulse generation. Operating the balloon at an angle to the horizontal, which is the preferred nursing position in intensive care units, results in reduced IAB inflation and deflation performance, possibly compromising its clinical benefits.
No takes yet. Share an insight, caveat, or question.
Bruti et al. (2015) studied Intra-aortic balloon pump counterpulsation (in vitro experimental mock circulation model). Intra-aortic balloon pump operation at a 45° angle vs. Intra-aortic balloon pump operation at a horizontal position (0°) was evaluated on Duration of balloon deflation (s) (35% increase in deflation duration (MD 0.108 s), p=<0.01). Operating the intra-aortic balloon at a 45° angle slowed deflation by 35% and reduced deflation pressure pulse by 14.7% compared to the horizontal position.