Key result
Elecsys STAT hs-cTnT assay demonstrates high precision and accuracy to validate ESC NSTEMI rule-out criteria.
Why the study?
ESC guidelines recommend dynamic cTn determination for NSTEMI rule-out, but cutoffs for discharge are lower than the limits of quantification proposed by reagent suppliers.
The Roche Elecsys STAT highly sensitive troponin T assay demonstrates acceptable precision, accuracy, and linearity at low concentrations, validating its use for the 0-1h rule-out algorithm in ESC guidelines for NSTEMI.
BACKGROUND: The European Society of Cardiology (ESC) guidelines recommend a dynamic (0-1h) cardiac troponin (cTn) determination for non-ST elevation myocardial infarction diagnosis. For patients with low cTn levels, a discharge from emergency can be considered. Nevertheless, cTn cutoffs for discharge are lower than the limits of quantification proposed by laboratory reagent suppliers. AIM: Validate cTn assay on the Elecsys STAT kit. MATERIALS AND METHODS: Precision, trueness, repeatability and within-laboratory variability were calculated from internal quality control and plasma pooled at 5.78 and 10.73 ng/L. Accuracy was calculated from external quality control. Uncertainty of measurement was calculated from (i) the uncertainty of the standard and control values and (ii) by precision from pooled plasma. Distribution of precision results from pooled plasma has been evaluated by bootstrap simulations. Dilution linearity tests with patient plasma were performed to evaluate the method for values near 5 ng/L. RESULTS: Precision and trueness ranged from 1.35 to 4.45% and from 0.14 to -3.74%, respectively. Accuracy results ranged from 101.40 to 104.90%. Within laboratory variability was 2.91%. Uncertainty ranged from 3.66% to 19.90% for higher (2188) to lower values (5.78 ng/L). Bootstrap simulations allowed utilization of precision data from pooled plasma to evaluate cTn assay. The method was linear from 4.48 to 39.80 ng/L. A linear regression model best described the data. CONCLUSION: Elecsys STAT method provides accurate cTn results, including patients with cTn results categorizing them as 'rule-out' in the ESC guidelines.
No takes yet. Share an insight, caveat, or question.
Metsu et al. (2023) studied Non-ST elevation myocardial infarction (NTSTEMI). Roche Elecsys STAT highly sensitive troponin T assay was evaluated on Assay precision, trueness, accuracy, and linearity. The Elecsys STAT highly sensitive troponin T assay demonstrated high precision (1.35-4.45%) and accuracy (101.4-104.9%), validating its use for the ESC guidelines 'rule-out' criteria for NTSTEMI.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: