We describe the development of a field-portable colorimetric immunoassay for the detection of Escherichia coli O157:H7, using antibody-directed liposomes (immunoliposomes) encapsulating dye as an analytical reagent. Antibodies (anti-E. coli O157:H7) thiolated by 2-iminothiolane were coupled to malemide-tagged liposomes encapsulating the marker dye, sulforhodamine B. Transmission electron microscopy showed that the immunoliposomes bound only to the serotype without any cross-reactivity with tested negative controls. A wicking reagent containing immunoliposomes and the test sample and a plastic-backed nitrocellulose strip with a measurement zone were used in a sandwich (noncompetitive) assay format. During the capillary migration of the wicking reagent, E. coli, with surface-bound immunoliposomes, was captured at the measurement zone on which antibodies to E. coli O157:H7 were immobilized. The color density of the measurement zone was directly proportional to the amount of E. coli O157:H7 in the sample. The detection limit of the current assay with pure cultures of the serotype was ca. 10(4) colony-forming units (CFU)/mL. The assay, which does not need washing and incubation steps, can be completed in 8 min. These results demonstrate the feasibility of using dye-encapsulating immunoliposomes in microporous membranes for the rapid detection of molecules with multivalent antigenic sites.
No takes yet. Share an insight, caveat, or question.
Park et al. (2000) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: