We report on a systematic study of time resolution made with three different commercial silicon photomultipliers (SiPMs) (Hamamatsu MPPC S10931-025P, S10931-050P, and S10931-100P) and two LSO scintillating crystals. This study aimed to determine the optimum detector conditions for highest time resolution in a prospective time-of-flight positron emission tomography (TOF-PET) system. Measurements were based on the time over threshold method in a coincidence setup using the ultrafast amplifier-discriminator NINO and a fast oscilloscope. Our tests with the three SiPMs of the same area but of different SPAD sizes and fill factors led to best results with the Hamamatsu type of 50×50×μm2single-pixel size. For this type of SiPM and under realistic geometrical PET scanner conditions, i.e., with 2×2×10×mm3LSO crystals, a coincidence time resolution of 220 ±4 ps FWHM could be achieved. The results are interpreted in terms of SiPM photon detection efficiency (PDE), dark noise, and photon yield.
No takes yet. Share an insight, caveat, or question.
Gundacker et al. (2012) studied this question.