Key result
Contour sound spectrograms detected aortic ball variance, showing peak frequencies <1,300 cycles/sec in 11 of 12 affected patients compared to >1,300 cycles/sec in 24 of 25 controls.
Why the study?
Does contour sound spectrography accurately detect aortic ball variance in patients with Starr-Edwards aortic prostheses?
Case-Control (n=37)
Does contour sound spectrography accurately detect aortic ball variance in patients with Starr-Edwards aortic prostheses?
Sound spectrography provides reliable objective evidence of aortic ball variance in patients with Starr-Edwards Model 1000 series prostheses by detecting decreased peak frequencies of the aortic opening sound.
Answers are grounded in this paper and cite their sources.
May support noninvasive detection of Starr-Edwards ball variance; leaves open prospective validation before any practice change.
Abnormalities in the silastic poppets of cardiac valve prostheses have been detected with increasing frequency. Ball variance can cause serious mechanical dysfunction of the prosthesis and can result in sudden death. Contour sound spectrograms were recorded on 12 patients with ball variance confirmed by operation and 25 controls. In ball variance the frequency of the aortic opening sound at the second right intercostal space was decreased, with the peak frequency recorded being below 1,300 cycles/sec in 11 of the 12 patients. The peak frequency was greater than 1,300 cycles/sec in 24 of the 25 control patients. The remaining patient had peak frequencies in both the normal and abnormal range. The diagnosis of aortic ball variance in patients with triple valve replacement remains difficult because of the nearly synchronous tricuspid closing and aortic opening sounds. The sound spectrographic findings have been the most reliable objective evidence of ball variance in patients with Starr-Edwards aortic prostheses of the Model 1000 series.
No takes yet. Share an insight, caveat, or question.
Hylen et al. (1969) conducted a case-control in Aortic ball variance (n=37). Aortic ball variance vs. Controls was evaluated on Peak frequency of the aortic opening sound below 1,300 cycles/sec. Contour sound spectrograms detected aortic ball variance, showing peak frequencies <1,300 cycles/sec in 11 of 12 affected patients compared to >1,300 cycles/sec in 24 of 25 controls.