Key result
Mesh nebulization of milrinone delivers ~30% higher absolute inhaled dose vs jet nebulization.
Why the study?
Inhaled milrinone reduces pulmonary hypertension during cardiac surgery, but its pharmacokinetics have not been determined for this route of administration.
Does mesh nebulization compared to jet nebulization increase the delivered dose and early plasma concentrations of inhaled milrinone in pulmonary hypertensive patients undergoing cardiac surgery?
RCT (n=12)
randomized
No
Does mesh nebulization compared to jet nebulization increase the delivered dose and early plasma concentrations of inhaled milrinone in pulmonary hypertensive patients undergoing cardiac surgery?
Mesh nebulization of inhaled milrinone delivers a significantly higher dose and achieves higher early plasma concentrations compared to jet nebulization in cardiac surgery patients, while remaining within the therapeutic range without causing systemic hypotension.
Answers are grounded in this paper and cite their sources.
Mesh nebulization may improve inhaled milrinone delivery efficiency; hypothesis-generating for clinical outcomes in right ventricular failure.
Inhaled milrinone administered before cardiopulmonary bypass (CPB) reduces the severity of pulmonary hypertension during cardiac surgery. However, milrinone pharmacokinetics has not been determined for this route of administration. The objective of this study was to investigate inhaled milrinone dosing in vitro and early plasma concentrations in vivo after jet and mesh nebulization. Twelve pulmonary hypertensive patients scheduled for cardiac surgery were randomized to receive milrinone (5 mg) by inhalation before CPB using a jet or mesh nebulizer. In vitro experiments were conducted to determine the inhaled dose delivered with either jet or mesh nebulization. In vivo experiments involved hemodynamic monitoring and blood samples drawn from patients for the first 15 min after the end of inhalation to determine early plasma concentrations. After mesh nebulization, the mean in vitro inhaled dose was almost 3-fold higher compared to jet nebulization (46.4% vs 16.6% for mesh and jet, respectively; mean difference, 29.8%; 95% CI, 14.1 to 45.5; P = 0.006). Consistent with this, the early plasma concentrations in vivo were also 2-3 fold higher after mesh nebulization (P = 0.002-0.005). After inhalation (jet or mesh nebulization), milrinone early plasma concentrations remained within the therapeutic range. No systemic hypotension was reported in our patients.
No takes yet. Share an insight, caveat, or question.
Nguyen et al. (2020) conducted an RCT in Pulmonary hypertension in cardiac surgery (n=12). Inhaled milrinone via mesh nebulization vs. Inhaled milrinone via jet nebulization was evaluated on In vitro inhaled dose (MD 29.8%, 95% CI 14.1 to 45.5, p=0.006). Mesh nebulization of inhaled milrinone yielded a significantly higher in vitro inhaled dose compared to jet nebulization (46.4% vs 16.6%, MD 29.8%), with correspondingly higher plasma concentrations.