Answers are grounded in this paper and cite their sources.
PURPOSE: To assess the feasibility of a polytetrafluoroethylene-collared (PTFE) endoluminal graft in the treatment of lower limb atherosclerosis. METHODS: We designed an endograft using a 3-mm-diameter balloon-expandable PTFE graft with terminal Palmaz stents placed on the outside of the graft, folding the PTFE back over the stents to form a collar at each end. Under protocol, this device was implanted in 8 symptomatic patients with lower limb ischemia. The lesions, ranging from 4 to 20 cm long, were located in the superficial femoral artery (n = 5), femoropopliteal segment (n = 1), and common (n = 1) and external (n = 1) iliac arteries. The device required a 14-F introducer system. RESULTS: Graft lengths varied from 4 to 35 cm. Implantation was successful in all cases, but procedural complications occurred in 4 patients (2 access site hematomas, 1 leading to endograft occlusion; 1 arterial injury, and 1 distal thromboembolism). At a mean 14-month follow-up, 5 endografts were patent (2 after reintervention for restenosis or thrombosis). The common iliac endograft and 2 superficial femoral artery devices occluded after 3, 2, and 12 months, respectively. CONCLUSIONS: Although this endoluminal graft system is technically feasible and showed encouraging intermediate-term patency in a small pilot study, the early and late complications identified several shortcomings of this design, which needs refinement.
No takes yet. Share an insight, caveat, or question.
Lundquist et al. (2000) studied this question.