Key result
SAVR with RESILIA tissue shows excellent early safety with ~94% two-year survival.
Why the study?
Does surgical aortic valve replacement with the Carpentier-Edwards PERIMOUNT Magna Ease aortic valve with RESILIA tissue provide safe and effective outcomes in patients requiring AVR?
Does surgical aortic valve replacement with the Carpentier-Edwards PERIMOUNT Magna Ease aortic valve with RESILIA tissue provide safe and effective outcomes in patients requiring AVR?
The novel RESILIA tissue bioprosthetic aortic valve demonstrates excellent early safety, hemodynamics, and clinical effectiveness at 2 years post-surgical aortic valve replacement.
Answers are grounded in this paper and cite their sources.
Supports early safety evaluation of RESILIA tissue in SAVR; leaves open durability and comparative efficacy questions.
OBJECTIVES: The COMMENCE trial was conducted to evaluate the safety and effectiveness of a novel bioprosthetic tissue for surgical aortic valve replacement (AVR). METHODS: Patients underwent clinically indicated surgical AVR with the Carpentier-Edwards PERIMOUNT™ Magna Ease™ aortic valve with RESILIA™ tissue (Model 11000A) in a prospective, multinational, multicentre (n = 27), single-arm, FDA Investigational Device Exemption trial. Events were adjudicated by an independent Clinical Events Committee; echocardiograms were analysed by an independent Core Laboratory. RESULTS: Between January 2013 and February 2016, 689 patients received the study valve. Mean age was 67.0 ± 11.6 years; 71.8% were male; 26.3% were New York Heart Association Class III/IV. Mean STS PROM was 2.0 ± 1.8 (0.3-17.5). Isolated AVR was performed in 59.1% of patients; others had additional concomitant procedures, usually CABG. Thirty-day outcomes for all patients included all-cause mortality 1.2%, thromboembolism 2.2%, bleeding 0.9%, major paravalvular leak 0.1% and permanent pacemaker implantation 4.7%. Median intensive care unit and hospital length of stay were 2 (range: 0.2-66) and 7 days (3.0-121.0), respectively. At 2 years, New York Heart Association class improved in 65.7%, effective orifice area was 1.6 ± 0.5 cm2; mean gradient was 10.1 ± 4.3 mmHg; and paravalvular leak was none/trivial in 94.5%, mild in 4.9%, moderate in 0.5% and severe in 0.0%. One-year actuarial freedom from all-cause mortality for isolated AVR and for all patients was 98.2% and 97.6%, respectively. Two-year actuarial freedom from mortality in these groups was 95.3% and 94.3%, respectively. CONCLUSIONS: These data demonstrate excellent early safety and effectiveness of aortic valve replacement with a novel bioprosthetic tissue (RESILIA™). CLINICAL TRIAL REGISTRATION: clinicaltrials.gov: NCT01757665.
No takes yet. Share an insight, caveat, or question.
Puskas et al. (2017) studied Surgical aortic valve replacement (n=689). Carpentier-Edwards PERIMOUNT Magna Ease aortic valve with RESILIA tissue was evaluated on 2-year actuarial freedom from mortality. Surgical aortic valve replacement with a novel bioprosthetic tissue (RESILIA) demonstrated excellent early safety, with a 2-year actuarial freedom from mortality of 94.3%.
Synapse has enriched 10 related papers. Consider them for comparative context: