Key result
Simplified pulsatile models match complete models for intra-stent flow but cut computational time ~80%.
For numerical simulation of intra-stent flows, the simplified pulsatile model offers a more computationally efficient alternative to the complete pulsatile model without sacrificing accuracy near the stent struts.
Answers are grounded in this paper and cite their sources.
May facilitate efficient stent-flow simulations; leaves open clinical translation and validation beyond near-strut regions.
We evaluate here three hemodynamic models used for the numerical simulation of bare and stented artery flows. We focus on two flow features responsible for intra-stent restenosis: the wall shear stress and the re-circulation lengths around a stent. The studied models are the Poiseuille profile, the simplified pulsatile profile and the complete pulsatile profile based on the analysis of Womersley. The flow rate of blood in a human left coronary artery is considered to compute the velocity profiles. "Ansys Fluent 14.5" is used to solve the Navier-Stokes and continuity equations. As expected our results show that the Poiseuille profile is questionable to simulate the complex flow dynamics involved in intra-stent restenosis. Both pulsatile models give similar results close to the strut but diverge far from it. However, the computational time for the complete pulsatile model is five times that of the simplified pulsatile model. Considering the additional "cost" for the complete model, we recommend using the simplified pulsatile model for future intra-stent flow simulations.
No takes yet. Share an insight, caveat, or question.
Chabi et al. (2015) studied Intra-stent restenosis. Simplified pulsatile profile and complete pulsatile profile vs. Poiseuille profile was evaluated on Wall shear stress and re-circulation lengths around a stent. The simplified pulsatile model provided similar results to the complete pulsatile model close to the strut but required one-fifth of the computational time for intra-stent flow simulations.
Synapse has enriched 10 related papers. Consider them for comparative context: