Key result
Endovascular snare technique facilitates successful SE-THV delivery in angulated aortas after conventional approaches fail.
Why the study?
Angulated aortas present an additional challenge during transcatheter aortic valve-in-valve replacement, particularly when using self-expanding transcatheter heart valves.
Does the endovascular snare technique facilitate successful delivery of self-expanding transcatheter heart valves during transcatheter aortic valve-in-valve replacement in patients with angulated aortas?
Case Report (n=3)
Does the endovascular snare technique facilitate successful delivery of self-expanding transcatheter heart valves during transcatheter aortic valve-in-valve replacement in patients with angulated aortas?
The endovascular snare technique is a feasible option for facilitating the delivery of self-expanding transcatheter heart valves during valve-in-valve procedures in patients with angulated aortas.
May aid bailout delivery in angulated aortas; hypothesis-generating and requires prospective validation before wider use.
BACKGROUND: Transcatheter aortic valve-in-valve replacement (ViV) has been widely accepted as a less invasive alternative to treat failed aortic surgical or transcatheter bioprosthetic valves. Angulated aortas present an additional challenge, particularly when using self-expanding transcatheter heart valves (SE-THV). METHODS: Two patients with failed surgical bioprosthetic aortic valves and one patient with a failed transcatheter bioprosthetic aortic valve underwent transcatheter aortic ViV using SE-THV. All were deemed high-risk for surgical aortic valve replacement by a heart team. All three patients had initial failed SE-THV delivery using a conventional approach with subsequent successful delivery using the endovascular snare technique. RESULTS: In Cases 1 and 2, the SE-THV was biased towards the greater curve of the angulated aorta and behind the outer frame of the bioprosthetic valve frame. An endovascular snare was deployed through a secondary left femoral artery access, and the valve delivery system was advanced through the snare in the ascending aorta. The snare was tightened around the SE-THV capsule proximal to the hat-marker, allowing deflection of the SE-THV and successful delivery. In Case 3, the SE-THV interacted with the tall frame of a failed SE-THV. A snare via the left femoral artery was deployed in the descending artery. The SE-THV was advanced through the snare, and both the snare and SE-THV were advanced together to the ascending aorta where the SE-THV was deflected and successfully delivered. CONCLUSIONS: The endovascular snare technique is a feasible option for successful delivery of SE-THV during transcatheter aortic ViV in failed transcatheter or surgical bioprosthetic valves in angulated aortas.
No takes yet. Share an insight, caveat, or question.
Rohm et al. (2021) conducted a case report in Failed bioprosthetic aortic valves in angulated aortas (n=3). Endovascular snare technique was evaluated on Successful delivery of SE-THV. The endovascular snare technique facilitated successful delivery of self-expanding transcatheter heart valves in 3 patients with angulated aortas after conventional delivery approaches failed.