Key result
In an in silico model, the inversion gap reduced the arterial spin labeling MRI signal to 64-74% in the sagittal plane and 53-84% in the coronal plane compared to the absence of a gap.
In silico modeling reveals that MRI ASL measurements of pulmonary blood flow are subject to significant errors depending on imaging plane location due to inversion gaps and conduit vessel contributions.
Answers are grounded in this paper and cite their sources.
May bias pulmonary ASL MRI blood flow estimates; hypothesis-generating for plane-specific corrections pending in vivo validation.
MRI images of pulmonary blood flow using arterial spin labeling (ASL) measure the delivery of magnetically tagged blood to an image plane during one systolic ejection period. However, the method potentially suffers from two problems, each of which may depend on the imaging plane location: 1) the inversion plane is thicker than the imaging plane, resulting in a gap that blood must cross to be detected in the image; and 2) ASL includes signal contributions from tagged blood in conduit vessels (arterial and venous). By using an in silico model of the pulmonary circulation we found the gap reduced the ASL signal to 64-74% of that in the absence of a gap in the sagittal plane and 53-84% in the coronal. The contribution of the conduit vessels varied markedly as a function of image plane ranging from ∼90% of the overall signal in image planes that encompass the central hilar vessels to <20% in peripheral image planes. A threshold cutoff removing voxels with intensities >35% of maximum reduced the conduit vessel contribution to the total ASL signal to ∼20% on average; however, planes with large contributions from conduit vessels underestimate acinar flow due to a high proportion of in-plane flow, making ASL measurements of perfusion impractical. In other image planes, perfusion dominated the resulting ASL images with good agreement between ASL and acinar flow. Similarly, heterogeneity of the ASL signal as measured by relative dispersion is a reliable measure of heterogeneity of the acinar flow distribution in the same image planes.
No takes yet. Share an insight, caveat, or question.
Burrowes et al. (2012) studied Pulmonary blood flow measurement. Arterial spin labeling (ASL) MRI was evaluated on ASL signal reduction. In an in silico model, the inversion gap reduced the arterial spin labeling MRI signal to 64-74% in the sagittal plane and 53-84% in the coronal plane compared to the absence of a gap.
Synapse has enriched 10 related papers. Consider them for comparative context: