Why the study?
Does FAIRER compared to FAIR ASL MRI improve visualization of pulmonary parenchyma perfusion in healthy volunteers?
Does FAIRER compared to FAIR ASL MRI improve visualization of pulmonary parenchyma perfusion in healthy volunteers?
Pulsed ASL techniques like FAIR and FAIRER can visualize pulmonary parenchyma perfusion with negligible motion artifacts.
Answers are grounded in this paper and cite their sources.
May support ASL MRI for pulmonary perfusion research in volunteers; leaves open clinical translation pending patient validation.
Magnetic resonance imaging of pulmonary parenchyma perfusion using pulsed arterial spin labeling (ASL) techniques is presented. ASL uses magnetically labeled water as an endogenous, freely diffusible tracer. Presented are comparative results of ASL methods called Flow sensitive Alternating Inversion Recovery (FAIR), and FAIR with an Extra Radiofrequency pulse (FAIRER). Six healthy human volunteers were imaged. Perfusion-weighted images at different time delays, TI, were calculated from the subtraction of the control and tag images, which were acquired within a single breathhold. Detailed pulmonary structures can be visualized with negligible cardiac or respiratory motion artifacts. Different patterns of signal enhancement between the pulmonary vessels and parenchyma are shown in the perfusion images acquired at different TIs.
No takes yet. Share an insight, caveat, or question.
M. et al. (1999) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: