Key result
Free breathing FAIR cardiac ASL without navigators is feasible, with a 4-second repetition time (TP2) offering the best alternative and detecting a stress-induced MBF increase of 0.22 ml/g/min (P=0.037).
Why the study?
Arterial spin labeling allows noninvasive tissue blood flow measurement, but human cardiac application remains challenging, prompting efforts to improve FAIR ASL scan efficiency.
Does a specific repetition time optimize free breathing myocardial arterial spin labeling in healthy volunteers?
Cross-Sectional (n=14)
Does a specific repetition time optimize free breathing myocardial arterial spin labeling in healthy volunteers?
p-value: p=0.037
Free breathing FAIR cardiac ASL without navigators is feasible, and a TR of approximately four seconds (TP2) offers the optimal balance of scan efficiency and signal quality.
Enables navigator-free free-breathing cardiac ASL in volunteers; leaves open clinical translation pending patient validation.
Arterial spin labeling (ASL) offers a noninvasive and repeatable measurement of tissue blood flow, but its application in the human heart is still challenging. The aim of this study was to improve the scan efficiency of flow‐sensitive alternating inversion recovery (FAIR) ASL using a single inversion time. First, by evaluating a free breathing acquisition without navigators. Second, by shortening TR to acquire more ASL pairs. A FAIR ASL with balanced steady state free precession (bSSFP) readout was implemented at 3 T and tested in 14 healthy volunteers. Three cardiac‐triggered ASL scans with different TRs were compared while maintaining a constant acquisition time. These scans were named by the trigger pulse (TP) that started each acquisition: every two (TP1), four (TP2) and six (TP3) cardiac cycles. Two studies were performed on different days to assess the reproducibility of TP2 and TP3 scans. Perfusion response during passive leg raising was also evaluated. Student t‐test and nonparametric equivalent were computed for perfusion differences. Bland–Altman and coefficient of variation (CV) were calculated for the reproducibility assessment. Mean myocardial blood flow (MBF) values measured at rest were: 1.47 ± 0.91 ml/g/min (TP1), 1.95 ± 0.93 ml/g/min (TP2) and 1.94 ± 1.41 ml/g/min (TP3). Intrasession CV was 17.15% (TP2) and 24.17% (TP3) and intersession CV was 36.86% (TP2) and 18.38% (TP3). Mean MBF was found to increase with passive stress by 0.22 ± 0.41 ml/g/min for TP2 ( P = 0.037). Mean MBF values were consistent with the literature, although in the high end of the normal range. The lower MBF values measured in TP1 scans are attributable to saturation of blood outside the slice because of shorter TR. Free breathing FAIR cardiac ASL without navigators is feasible. A TR of approximately four seconds (TP2) offers the best alternative with similar mean ASL and tSNR than longer TR acquisitions.
No takes yet. Share an insight, caveat, or question.
Aramendía‐Vidaurreta et al. (2019) conducted a cross-sectional in Healthy (n=14). Flow-sensitive alternating inversion recovery (FAIR) ASL with different repetition times vs. Comparison between trigger pulses every two (TP1), four (TP2), and six (TP3) cardiac cycles was evaluated on Mean myocardial blood flow (MBF) at rest and during passive stress (p=0.037). Free breathing FAIR cardiac ASL without navigators is feasible, with a 4-second repetition time (TP2) offering the best alternative and detecting a stress-induced MBF increase of 0.22 ml/g/min (P=0.037).